DOIONLINE

DOIONLINE NO - IJIEEE-IRAJ-DOI-8660

Publish In
International Journal of Industrial Electronics and Electrical Engineering (IJIEEE)-IJIEEE
Journal Home
Volume Issue
Issue
Volume-5,Issue-7  ( Jul, 2017 )
Paper Title
Real Time Blood Sugar Monitoring
Author Name
Dewiani, Elyaspalantei, Ika Puspita
Affilition
Electrical Engineering Department Hasanuddin University, South Sulawesi, INDONESIA
Pages
114-118
Abstract
This research aims to design an interface that connects the blood sugar measuring instrument with the mobile phone as well as designing database systems and visual display for monitoring blood sugar levels in a patient's body. Monitoring system that integrates glucometer, arduino, Bluetooth, and mobile phones, are used to transmit data from the blood sugar measurement. Beside the position of patient there is glucometer along with android based mobile phone that will be used to transmit the measured data to a web server that can be accessed by doctors to monitor the progress of patients' health. Based on observations, we obtained error value of 23.33% for observation within 60 seconds, an error of 32% for observation within 90 seconds, and the error of 31.11% for observation within 120 seconds. Keywords- Glucometer, Diabetes, Bluetooth, Handphone.
  View Paper