DOIONLINE

DOIONLINE NO - IJMPE-IRAJ-DOIONLNE-6429

Publish In
International Journal of Mechanical and Production Engineering (IJMPE)-IJMPE
Journal Home
Volume Issue
Issue
Volume-4,Issue-12  ( Dec, 2016 )
Paper Title
Comparison of Experimental Measurements and Finite Element Predictions of Contact Mechanics of Metal-on-Polyethylene Total Hip Replacement
Author Name
Ahmet C.Cilingir, Ahu Dogmus
Affilition
Mechanical Engineering Department,Sakarya University, Sakarya, Turkey
Pages
75-78
Abstract
Metal-on-ultra high molecular weight polyethylene (UHMWPE) total hip replacement (THR) has been the most popular and clinically successful hip prosthesis to date. The long-term performance of THR depends on both the tribological characteristics and biomechanical behaviour of the prosthesis. Experimental approaches have the limitations of high costs, time-consuming and complex. Therefore, a simplified finite element model was developed to compare the measured contact areas and contact pressures from experimental model. The contact areas on the articulating surface of polyethylene cup under different loads were measured experimentally in self-made joint simulator and then maximum contact pressures were calculated from these measurements according to Hertzian Contact Theory. Good agreements of contact areas and contact pressures between finite element predictions from the simplified model and experimental measurements and calculations were obtained. This indicated that the simplifications and assumptions made for the finite element model were reasonable and finite element predictions from the simplified model were valid. Keywords- Total Hip Replacement, Metal-on-Polyethylene, Contact Mechanics.
  View Paper