DOIONLINE

DOIONLINE NO - IJACSCC-IRAJ-DOIONLINE-14202

Publish In
International Journal of Advances in Computer Science and Cloud Computing (IJACSCC)-IJACSCC
Journal Home
Volume Issue
Issue
Volume-6,Issue-2  ( Nov, 2018 )
Paper Title
Linac Quality Assurance Software
Author Name
Aleksei E. Zhdanov, Leonid G. Dorosinsky
Affilition
Ural Federal University named after the first President of Russia B.N.Yeltsin, Russia
Pages
1-6
Abstract
For radiotherapy, quality assurance is necessary to provide precise and accurate radiation treatment. Besides treatment plans and patient positioning, machine status is another vital issue that affects radiotherapy. In this article, we perform Linear Accelerator (Linac) quality assurance (QA) software based on MATLAB. All dates and images which were used in this article were taken by Linac of radiation therapy department of University Clinic of Erlangen. We have tested the accuracy of beam isocenter, conformity of light field and irradiation field, multileaf collimator (MLC) leakage and dynamic MLC positions relative to rotation, and in this article we will include the results of these tests and analyze machine status by parameters mentioned above. Finally, we will discuss about the validity of existing machine quality assurance program. Keywords - linear accelerator, quality assurance, image processing, multileaf collimator, light field, irradiation field
  View Paper